| For increased power calibration The Process Calibrator Fluke 726 Precision Multifunction designed specifically for process applications to ensure a high working ability, great power and precision calibration foolproof. The Fluke 726 can measure and all process parameters, but also calibrate almost anything. In addition, it interprets the results of calibration without the need for computer and can store measurement data for later analysis. |

- More accurate and better performance for generating calibration accuracy 0.01%
- Two independent channels, measurement, generation and simultaneous display of process signals
- Measure volts, mA, RTDs, thermocouples, frequency and resistance to test sensors and transmitters
- Source / simulate volts, mA, thermocouples, RTDs, frequency and pressure to calibrate transmitters
- * Measure or pressure by any of 29 Fluke pressure modules 700Pxx
- Generation of current (mA) with simultaneous pressure measurement to conduct valve and tests / P
- Test pressure switch built to capture the setting, reset and deadband of a switch
- Possibility of storage of 8 calibration results for later analysis
- Calculating percentage error of the transmitter and interpretation of calibration resultswithout computer
- Running fast linearity tests with features of stepper automatic and automatic ramp
- Power transmitters during test (24 V) with simultaneous mA measurement
- Storing frequently used test setups for later use
- Frequency counter and method of generation of pulse trains of frequency for more precise control of flow
- HART mode adding a resistance of 250 ohms measurement and generation of mA to ensure full compatibility with HART
- Customizing curves RTDs, add calibration constants for probes RTDs certified in order to obtain more accurate temperature measurements
- New design incorporating a voltage input protection for increased reliability
- 3 years warranty
|
* Pump pressure required |
recision measuring | Voltage DC | 30.000 V | 0.02% + 2 points (upper display) | 30.000 V | 0.02% + 2 points (lower display) | 100.00 mV | 0.01% + 2 points | of -10.00 mV to 75.00 mV | 0.025% + 1 point (via TC connector) |
| DC | 24.000 mA | 0.01% + 2 points |
| Resistance | from 0.0 Ω to 4 000.0 Ω | + / - 0.015% |
| Frequency | 2.0 1 000.0 CPM CPM | 0.05% + 1 point | from 1.0 Hz to 1 000.0 Hz | 0.05% + 0.1 Hz | from 1.0 kHz to 15.0 kHz | 0.05% + 0.1 kHz | from 10.0 kHz to 15.0 kHz | 0.05% + 1 point | Sensitivity | 1 Vpp minimum |
| Pressure | Accuracy | of 0.025% of range using any of 29 pressure modules. | | (For detailed specifications, refer to pressure modules in the Options and accessories) | | Modules available for differential, gage, absolute, double and high, and the void. |
|
Source Accuracy | Voltage DC | 100.00 mV | 0.01% + 2 points | 10.000 V | 0.01% + 2 points | of -10.00 mV to 75.00 mV | 0.025% + 1 point (via TC connector) |
| DC | 24.000 mA (Gen.) | 0.01% + 2 points | 24.000 mA (simulation) | 0.01% + 2 points |
| Resistance | Range | from 5 Ω to 400 Ω | | from 5 Ω to 400 Ω | | from 400 Ω to 1500 Ω | | of 1500 Ω to 4000 Ω | Accuracy | Percentage of reading + floor | | 0.015% + 0.1 Ω | | 0.015% + 0.05 Ω | | 0.015% + 0.5 Ω | | 0.015% + 0.5 Ω | Excitation current | 0.1 mA to 0.5 mA | | 0.5 mA to 3 mA | | 0.05 mA to 0.8 mA | | 0.05 mA to 0.4 mA |
| Frequency | 2.0 1 000.0 CPM CPM | 0.05% | from 1.0 Hz to 1 100.0 Hz | 0.05% | of 1.00 kHz to 10.00 kHz | 0.25% | from 10.0 kHz to 15.0 kHz | 0.5% | Waveform: | square wave 5 Vpp, offset by -0.1 V |
|
RTDs and thermocouples | Precision measurement and generation: | Cu10 | 1.8 ° C | NI-120 | 0.15 ° C | PT-100 (385) | 0.25 ° C | PT-100 (392) | 0.15 ° C | PT-100 (391-JIS) | 0.25 ° C | PT-200 (385) | 0.75 ° C | PT-500 (385) | 0.35 ° C | PT-1000 (385) | 0.15 ° C | Resolution: | 0.01 ° C | J: | 0.4 ° C | K: | 0.5 ° C | T: | 0.4 ° C | E: | 0.4 ° C | A: | 1.4 ° C | S: | 1.4 ° C | B: | 1.4 ° C | L: | 0.45 ° C | U: | 0.45 ° C | N: | 0.6 ° C | XK: | 0.4 ° C | BP: | 1.1 ° C | Resolution | J, K, T, E, L, N, U: 0.1 ° C, 0.1 ° F, B, R, S: 1 ° C, 0.1 ° F |
|
Specifications | Ramp functions | Generation functions: | Voltage, current, resistance, frequency, temperature | Ramps | Slow ramp, fast ramp, levels of 25% |
| Loop Power Function | Voltage: | 24 V | Accuracy: | 10% | Maximum current: | 22 mA, protected against short circuits |
| Step Functions | Generation functions: | Voltage, current, resistance, frequency, temperature | Levels | 25% of the range, 100% of range |
|
Environmental Specifications | Operating Temperature | | Storage temperature | | Operating Altitude | |
Safety Features | Compliance with safety standards | | Certified EN 61010-1:1993, ANSI / ISA S82.01-1994, CAN / CSA C22.2 No. 1010.1:1992 |
|
Mechanical & General Specifications | Dimensions | | Weight | | Batteries | | Battery Replacement | | Separate battery compartment, accessible without breaking calibration seal |
| Side port connections | | Pressure module connector, also used for programming real-time remote |
|
|
|
|