The Fluke 983 can perform several tasks:
- Measuring the effectiveness of filters
- Control in industrial clean rooms
- Make an initial analysis of the quality of indoor air for effective collaboration with experts in IAQ
- Isolate the sources of particulate matter in the context of a sanitation
- Prepare a report on the effectiveness of repairs to the customer
- Sell new services from the demonstration of the need for maintenance and repair
With its extended function data recording and display its six lanes, the Fluke 983 allows the user to quickly perform a greater number of tests, without wasting time searching through data. The Fluke 983 is lightweight and easy to use. It has a comfortable strap and a durable rubber holster.
Simplified and accelerated analysis of air quality
The Fluke 983 particle counter measures and displays six channels simultaneously allocated based on the particle size, temperature and humidity. Compact, lightweight and autonomous, it allows easy handling with one hand. With the Fluke 983,you can take accurate measurements anywhere. With a rechargeable battery life of 8 hours and the extended recording function of 5000 samples, you can perform fast and comprehensive measures of air quality. The Fluke 983 is the ideal tool to determine the size or origin of the particles in the air.
- Simultaneous measurement and display of 6 lanes divided according to particle size, temperature and humidity
- Measures particles from 0.3 microns
- Sampling time selectable, encrypted data, programmable delay
- 5000 record date, time, counting, relative humidity, temperature, volume, alarms and marking
- Transfer recorded data to a PC using the supplied software
- Compact, standalone for easy handling with one hand
- Friendly and intuitive user interface
- Constant availability
- LCD backlight for use in all lighting conditions
- NiMH rechargeable battery with a life of 8 hours
- Protective Cover
|
Specifications |
Routes | | 0,3 - 0.5 to 1.0 - 2.0 - 5.0 to 10.0 microns |
|
Flow | | 0.1 cfm (2.83 L / min) controlled by an internal pump |
|
Counting modes | | Concentration, addition, audio |
|
Counting efficiency | | 50% at 0.3 microns, 100% for particles> 0.45 microns (by JIS B9921: 1997) |
|
Count Zero | | 1 count / 5 minutes (JIS B9921: 1997) |
|
Loss coincidences | | 5% at 2,000,000 particles per ft 3 |
|
Relative Humidity | | ± 7%, 20% to 90% noncondensing |
|
Temperature | | ± 3 ° C, 10 ° C to 40 ° C |
|
Data Storage | | 5000 record (FIFO) date, time, counts, relative humidity, temperature, volume, alarms and marking |
|
Alarms | | Counts, low battery, faulty sensor |
|
Delay | |
Input samples | |
Interface | | RS-232 and RS-485 via RJ-45 |
|
Environmental Specifications |
Operation: | | 10 ° C to 40 ° C, 20% to 90% relative humidity, noncondensing |
|
Storage: | | -10 ° C to 50 ° C, up to 90% relative humidity, noncondensing |
|
Mechanical & General Specifications |
Size | | 11.43 x 20.96 x 5.72 cm | | Dimensions (W x L x D) |
|
Weight | |
Power | | AC adapter, 12 V DC 2.5 A, 90 to 250 V AC, 50-60 Hz |
|
Rechargeable Battery | | NiMH, 4.8 V at 4.5 Ah; removable |
|
Life of the battery | |
Charging time | |
Calibration | | Polystyrene latex particles in air (NIST) |
|